Apnea Testing for the Determination of Brain Death in Patients Supported by Extracorporeal Membrane Oxygenation
Figures



Tables
| Circuit flow | 50 - 80 mL/kg/min |
| Sweep gas flow | 50 - 80 mL/kg/min |
| Fractional inspired oxygen | 100% |
| Oxygen saturation (drainage cannula) | > 65% |
| Oxygen saturation (return cannula) | 100% |
| Arterial oxygen saturation | VA: > 95%; VV: 85-92% |
| Arterial carbon dioxide tension | 35 - 45 mm Hg |
| Mean arterial pressure | 65 - 95 mm Hg |
| Report | Case | Initial ECMO settings | Oxygenation method | ECMO during testing | Ancillary testing |
|---|---|---|---|---|---|
| Jarrah et al [21] | #1 | Blood flow rate 88 mL/kg/min Sweep flow rate 0.5 L/min FiO2 100% | CPAP FiO2 100% | Blood flow rate 88 mL/kg/min Sweep flow rate 0.1 L/min FiO2 100% | None |
| #2 | Blood flow rate 38 mL/kg/min Sweep flow rate 5.5 L/min FiO2 100% | CPAP FiO2 100% | Blood flow rate 38 mL/kg/min Sweep flow rate 1.0 L/min FiO2 100% | None | |
| #3 | Blood flow rate 140 mL/kg/min Sweep flow rate 0.22 L/min FiO2 100% | CPAP FiO2 100% | Blood flow rate 140 mL/kg/min Sweep flow rate 0.1 L/min FiO2 100% | EEG | |
| Hoskote et al [24] | Blood flow rate 4.5 mL/kg/min Sweep flow rate 6 L/min FiO2 100% | None | Blood flow rate 4.5 mL/kg/min Sweep flow rate 0.5 L/min FiO2 100% | None | |
| Smilevitch et al [22] | Blood flow rate 4.1 mL/kg/min Sweep flow rate 3 L/min FiO2 40% increased to 100% during preoxygenation | T-piece 9 L/min | Blood flow rate 4.1 mL/kg/min Sweep flow rate 3 L/min FiO2 100% | EEG | |
| Goswami et al [25] | #1 | Blood flow rate 5.0 mL/kg/min Sweep flow rate Not documented FiO2 100% | Not specified | Blood flow rate 5.0 mL/kg/min Sweep flow rate 0.5 L/min FiO2 100% | EEG |
| #2 | Blood flow rate 4.1 mL/kg/min Sweep flow rate 3 L/min FiO2 40% increased to 100% during preoxygenation | Not specified | Blood flow rate 6.0 mL/kg/min Sweep flow rate 0.5 L/min FiO2 100% | EEG | |
| Iannuzzi et al [23] | Blood flow rate 4.0 mL/kg/min Sweep flow rate 8 L/min FiO2 70% | CPAP FiO2 100% | Blood flow rate 2.0 mL/kg/min Sweep flow rate 0.5 L/min FiO2 100% | None |